AC Hipot | |
---|---|
Frequency | 50 or 60Hz selectable |
Output Voltage | 100 – 5000v |
Output Voltage Setting | 10V per step |
Measurement Display Range | 0.001mA – 20.000mA |
Current Display Resolution | 0.001mA |
Max Rated Load | 100VA |
Arc Detection Current | 1mA - 40mA |
Insulation Resistance | |
---|---|
Output Voltage | 50 – 5000V |
Output Voltage Setting | 10V per Step |
Measurement when Test Voltage (500V – 5000V DC) | |
---|---|
Display Range | 1.00GΩ - 50.00GΩ |
Resolution | 10MΩ |
Ground Bond | |
---|---|
Output Current Frequency | 50Hz or 60Hz Selectable |
Output Test Current | 1.00A to 40.00Arms |
Test Current Resolution | 0.01A |
Resistance Measuring Range | 0.000Ω to 1.999Ω |
Resolution | 0.001Ω |
Hold Time (Hipot and Insulation Resistance Tests) | |
---|---|
Hold Time (Hipot & IRT) Duration | 0.1s – 300s |
Hold Time (Hipot & IRT) Resolution | 0.1s |
Hold Time (Ground Bond Test) | |
---|---|
Hold Time (GBT) Duration | 0.5s – 240s |
Hold Time (GBT) Resolution | 0.1s |
Power Requirements | |
---|---|
Mains Supply Voltage | 80V to 264V AC |
Mains Supply Frequency | 50/60Hz |
DC Hipot | |
---|---|
Output Voltage | 100 – 6000V |
Output Voltage Setting | 10V per Step |
Measurement Display Range | 0.001mA – 10.000mA |
Display Resolution | 0.001mA |
Arc Detection Current | 1mA - 40mA |
Measurement when Test Voltage (50V – 1000V DC) | |
---|---|
Display Range 1 | 0.01MΩ - 9.99MΩ |
DR1 Resolution | 0.01MΩ |
Display Range 2 | 10MΩ - 999MΩ |
DR2 Resolution | 1MΩ |
Continuity | |
---|---|
Test Method | DC 4 Wire Measurement |
Test Current | 250mA ± 30mA |
Resistance Display Range 1 | 0.00Ω to 9.99Ω |
DR1 Resolution | 0.01Ω |
Resistance Display Range 2 | 10.0Ω to 99.9Ω |
DR2 Resolution | 0.1Ω |
Resistance Display Range 3 | 100Ω to 999Ω |
DR3 Resolution | 1Ω |
Resistance Display Range 4 | 1.00kΩ to 9.99kΩ |
DR4 Resolution | 0.01kΩ |
Ramp Up and Down (All Tests) | |
---|---|
Ramp Up and Down Duration | 0.0s – 300s |
Ramp Up and Down Resolution | 0.1s |
Hold Time (Continuity Test) | |
---|---|
Hold Time (CT) Duration | 0.5s – 300s |
Hold Time (CT) Resolution | 0.1s |