Insulation Resistance | |
---|---|
Test Voltage | 250/500VDC @1mA nominal, < 2mA (s/c) |
Test Duration (Class I or II) | 2s |
Display Range | 0.01 Mohm to 19.99 Mohm |
Pass Value | 1.0 MΩ Class I, 2.0 MΩ Class II |
IEC Lead Test | |
---|---|
Test Voltage | 5V DC nominal |
Test Duration | Time for test (2s max) |
Test | Live / Neutral Checks for o/c, s/c and crossed |
Touch Current | |
---|---|
Test Voltage | Supply Voltage, max current 13A |
Test Duration | 30s maximum |
Display Range | 0.10 mA to 5.00 mA |
Pass Value | 0.25mA |
RCD trip time | |
---|---|
Test Voltage | Mains supply voltage |
Test Current | 10mA / 30mA rms sinusoidal |
Test Current Phase | 0°/180° auto switched on alternate tests |
Measuring Range | 0ms – 500ms |
Pass Value | 200ms |
Earth Continuity | |
---|---|
Output Current | 200mA minimum into 2Ω |
Test Voltage (o/c) | 9V DC nominal |
Test Duration (Class I Test) | 2s nominal |
Display Range | 0.01 ohm to 19.99 ohm |
Pass Value (Class I Test) | 0.2Ω |
Equivalent Leakage | |
---|---|
Test Voltage | 40V Nominal AC 50Hz Open Circuit |
Test Duration | 2s – 30s nominal |
Display Range | 0.1 mA to 19.99mA |
Pass Value | 0.75mA CLI / 0.25mA CL II |
Protective Conductor Current | |
---|---|
Test Voltage | Supply Voltage, max current 13A |
Test Duration | 30s maximum |
Display Range | 0.15 mA to 9.99 mA |
Pass Value | 0.75mA/3.5mA (user selectable with 0.75mA as default) |
3 phase differential leakage current (*Requires 3 phase adaptor accessory) | |
---|---|
Test Voltage | 3 phase supply voltage |
Test Duration | Continuous measurement |
Display Range | 0.10 mA to 9.99 mA |
Pass Value | 3.5mA |
Power Socket Check | |
---|---|
Input voltage range | 225 – 254V AC max current 13A UK |
Indicates configuration of voltage potential | Line potential phase to earth, Line potential phase to neutral, Line potential neutral to earth |